vtec
Set or read TEC voltage in V, where supported.
Usage
>> vtec
0.00000
>> vtec 1.2
1.20000
Product behavior
| Product | Access | Write arguments | Read response | Notes |
|---|---|---|---|---|
| CTL20, CTL200, CTL300E, CTL300P | Advanced write | vtec <voltage> |
monitored TEC voltage in V | Writes are accepted only when tecon 0 has disabled closed-loop TEC control. CTL200 enables the TEC supply before applying a valid manual voltage. CTL300P enables the VSTEC output before applying a valid manual voltage. |
| TEC200 | Advanced write | vtec <voltage> |
applied H-bridge TEC voltage in V | Writes are accepted only when tecon 0 has disabled closed-loop TEC control. The requested voltage is clamped to the current vtmin and vtmax limits before it is applied. |
| HTC200 | Read-only | none | monitored TEC/heater voltage in V | The write form is not available. |
Notes
On products that allow manual voltage writes, use vtmin and vtmax to review the configured voltage protection range before writing vtec. A successful write can immediately change the TEC drive voltage.